Specification | ||
Temperature range |
-10℃ ~ 60℃
|
||
Emissive area |
Φ300mm
|
||
Temperature resolution |
0.1℃ / 0.01℃
|
||
Accuracy |
±(0.3%FS+1.0)℃
/ ±(0.3%FS+0.5)℃(Ambient temp. 23℃±5 ℃)
|
||
Stability |
±(0.1 ~ 0.5)℃/10min / ±(0.01 ~ 0.05)℃/10min
|
||
Uniformity |
±0.5℃ (at 23℃±5℃ within 2/3 of the central area)
|
||
Emissivity |
0.95±0.02
|
||
Power supply |
220VAC 50Hz
|
||
Dimension/Weight |
Blackbody: W685mm×H595mm×D325mm
/ 54kg
|
||
Controller: W470mm×H245mm×D465mm / 23kg
|
|||
Operating ambient temp |
0℃ ~ 40℃(≤20℃ when blackbody operating in cooling mode)/ ≤ 80%RH
|
||
Option |
RS232 or RS485
communication interface/ Plexiglass cover
|
||
Extended type |
Operating ambient temp. range of the blackbody is -40℃~60℃
|
||
Product features:The applications of these sources are calibrations and uniformity test of IR thermometers, IR sensors, thermal imagers, etc.
Note 1:Kelvin provides the following solutions to avoid ice buildup on the radiation surface when blackbody operating at temperature below the dew point:
Install a plexiglass cover at the front panel of the blackbody, fill the cover with nitrogen or dry air (by customer), and seal the blackbody radiation surface with dry gas. Or the blackbody and the measured instrument are placed in a cabinet filled with dry gas.
Note 2: The inspection basis shall refer to JJF 1552-2015 Calibration Specification for Blackbody Radiation Sources of Radiation Thermometry from -10℃ to 200℃. |
· We reserve the right to change product technology and instructions at any time, products shall be subject to any changes without additional notices.