JQ-80MYD1B/JQ-80MYD1C Low-temperature Blackbody
Specification
Temperature range
-20℃ ~ 60℃
Emissive area
Φ80mm
Temperature resolution
0.1℃ / 0.01℃
Accuracy
±(0.3%FS+0.1)℃ / ±(0.3%FS+0.50)℃
Stability
±(0.1~0.3)℃/h / ±(0.01 ~ 0.03)℃/h
Emissivity
0.97±0.02
Power  supply
220VAC 50Hz 350W
Dimension/Weight
W235mm×H375mm×D295mm/14kg
Operating ambient temp
0℃ ~ 40℃(20 when operating in low temperature mode)/≤80%RH
Option
RS485 communication/Nitrogen inlet
Product features:The applications of these sources are calibrations of IR thermometers, IR sensors, thermal imagers, etc.

· We reserve the right to change product technology and instructions at any time, products shall be subject to any changes without additional notices.